Product information
- Material: B-D test paper, laminated sheets, laminated sheets with backing, foam sheets, and a sheet of cellophane wrapped around it, which is then sealed and packaged with an instruction label.
- Usage: determining the effect after the installation and commissioning of new sterilizers, and measuring the performance after the maintenance of sterilizers.
- Application: Suitable for Medical, Dental, Eye Center, Laboratory, Surgical Center, Pharmaceutical, and Pet Hospital, Biotechnology industry, Biomedical field; zoology research institute; biological laboratory.
- Pre-assembled disposable Bowie-Dick test pack for detection of defects in the sterilization process
- Easy-to-read test sheet with detailed instructions for use and interpretations of results
- Complies with ISO 11140-1, EN 867-4 and ISO 11140-4
Sterilization Process
The pattern on the B-D test paper will turn into a uniform dark brown to black when the sterilizer operates at 132℃ ~134℃ for 3.5 minutes. If the cold air is not completely expelled, the cold air inside the package will affect the penetration of the steam, resulting in a lighter-colored area in the center of the B-D test paper that is different from the surrounding areas.
Notes for Attention
- The test patterns inside the BD test kit must not come into contact with acidic or alkaline substances during storage and use.
- Avoid moisture (relative humidity should be less than 50%), store away from light, and protect from ultraviolet fluorescent lamps and sunlight.
- The test shall be conducted under steam conditions at 132~134℃ for no more than 4 minutes. The test shall be conducted in an empty autoclave before the first sterilization of the day.
- This product cannot be used to test the effectiveness of pressure steam sterilization.